KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG DISTRACTION, EXTERNAL; DISTRACTOR Back to Search Results
Model Number 51-601-00-09
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Impaired Healing (2378); Patient Problem/Medical Problem (2688)
Date of Event 06/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.Multiple mdr report were filed for this event, please see associated report: mdr 9610905-2018-00121.
 
Event or Problem Description
It was reported the patient fell and the distractor disengaged from the bone.Device was removed and replaced.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number was not identified therefore the device history records were not reviewed.The failure root cause cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTRACTION, EXTERNAL
Common Device Name
DISTRACTOR
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau 78570
GM  78570
MDR Report Key7630383
Report Number9610905-2018-00098
Device Sequence Number11217103
Product Code MQN
UDI-Device Identifier00888118083029
UDI-Public(01)00888118083029
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K034027
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
Report Date (Section F) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number51-601-00-09
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/11/2018
Event Location Hospital
Date Report to Manufacturer06/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date06/22/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age2 YR
-
-