• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; STEM, HUMERAL, LONG, REVISION, RSP, SIZE6

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; STEM, HUMERAL, LONG, REVISION, RSP, SIZE6 Back to Search Results
Model Number 507-06-175
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/11/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the stem loosening.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to loosening.The previous surgery and the revision detailed in this investigation occurred over 11 years and 8 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
STEM, HUMERAL, LONG, REVISION, RSP, SIZE6
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7630419
Report Number1644408-2018-00604
Device Sequence Number1898716
Product Code PHX
UDI-Device Identifier00888912144070
UDI-Public(01)00888912144070
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K041066
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2012
Device Model Number507-06-175
Device Catalogue Number507-06-175
Device Lot Number53800946
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/11/2018
Supplement Date Received by Manufacturer06/28/2018
Initial Report FDA Received Date06/22/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-00-000, LOT 53805572; 508-00-032, LOT 53799308; 508-32-103, LOT 53799319; 508-00-000, LOT 53805572; 508-00-032, LOT 53799308; 508-32-103, LOT 53799319
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
-
-