| Model Number |
89000 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts to obtain patient information have been unsuccessful.Patient information was declined by the facility.Further information provided, physical damage was observed when removing the device from the patient.There was no embolization, no injury and no decline in the status of the patient.The manufacturers device and (another manufacturers) guider were removed as a system.A second of the same make and model of the manufacturers device were used to complete the procedure.No tests/laboratory data was available.No physical product investigation was possible.To date, the device has not been returned to the manufacturer for analysis.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This complaint will be monitored as part of complaint data analysis.
|
| |
|
Event or Problem Description
|
|
This case was investigated according to the manufacturer's policy.It was reported, the catheter broke into half during removal.The broken half of the catheter remained in the guider and both were removed.No troubleshooting was performed.Nothing was left in the patient.This event is being reported because the device was removed as a system.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference: (b)(4).Visual and microscopic inspection was performed.The device was received in two pieces with a shaft separation distal to the telescope strain relief.The shaft separation is at the marker band.The marker band is applied with heat over a solid shaft as a single piece.Abrasions were observed on the marker band at the shaft separation location.The probable cause of the reported shaft separation failure most likely occurred during bending or tension on the catheter.Device integrity can be affected by external factors such as device manipulation, impact, and applied pressure associated with use.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
This case was reviewed and investigated according to the manufacturer's policy.This supplemental report (#1) is being submitted because the device was received and analyzed by the manufacturer.
|
| |
|