PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the system pcb assembly and completed other unrelated repairs.Proper device operation was observed through functional and performance testing and the device was returned to the customer for use.Physio further evaluated the removed system pcb assembly and determined that cause of the reported issue was due to integrated circuit, designator u61.U61 was electrically shorted and caused the device to lock up with a white screen and not complete the boot-up process.
 
Event Description
The customer contacted physio-control to report that the service light is illuminated on their device.There was no patient use associated with the reported issue.Upon evaluation of the customer's device, physio-control observed that the device would not boot up past the self-test screen.As a result, defibrillation may not be possible, if it were necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7630981
MDR Text Key112178798
Report Number0003015876-2018-00998
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Date Manufacturer Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-