|
SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 1-2 RT 10MM; PRSTHSS,KNE,PTLLFMRTBL,SMICNSTRND,CMNTD,PLYMRMTLPLYMR
|
Back to Search Results |
|
| Catalog Number |
74027212 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed due to insert disassociation from baseplate.The surgeon does not fault the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned.The clinical/medical team concluded, no clinical relevant documents were provided to conduct a thorough medical assessment. no medical assessment is warranted at this time. this complaint will be re-evaluated if more information becomes available.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batches/failure mode.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|