Model Number 15
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was not returned to physio-control for evaluation.A third-party service agent evaluated the device and verified the reported issue.After replacing the system pcb assembly, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
 
Event Description
The customer contacted physio-control to report that their device would not power on.When the on/off button was pressed the leds would illuminate and the display would flash.This would occur three times before the device would completely power off.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7631567
MDR Text Key112495458
Report Number0003015876-2018-01003
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001399
Was Device Available for Evaluation? No
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse