• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4004C0720
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the embolization coil stretch resistant wire (sr wire) was fractured and the proximal constraint sphere was missing.Conclusions: evaluation of the returned pc400 revealed that the sr wire was fractured.If the device is forcefully retracted against resistance, damage such as an sr wire fractured may occur.Further evaluation revealed that the non-penumbra microcatheter was kinked.The root cause of this kink could not be determined, however, this kink may have contributed to the reported resistance.The pusher assembly to the pc400 was not returned to penumbra for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400s).During the procedure, the physician successfully placed multiple pc400s and other coils using a non-penumbra microcatheter.The physician then felt resistance and was unable to advance another pc400 through the middle of the microcatheter and, therefore, the physician attempted to remove the pc400.While retracting, the pc400 unintentionally detached within the microcatheter.The physician then removed the microcatheter with the pc400 inside.After removal, the physician pushed the embolization coil out of the microcatheter to confirm that it was not implanted in the patient.The procedure was completed using a new microcatheter, new pc400s, and additional coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7631719
Report Number3005168196-2018-01239
Device Sequence Number14081462
Product Code HCG
UDI-Device Identifier00814548010915
UDI-Public00814548010915
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/16/2024
Device Catalogue Number4004C0720
Device Lot NumberF72322
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2018
Initial Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-