| Model Number |
BA28-80/I16-40 |
| Medical Device Problem Codes |
Leak/Splash (1354); Patient-Device Incompatibility (2682); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/29/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient was implanted with an afx bifurcated and limb stent in 2014.Recently, during a follow up scan the physician noted s calcified shelf n the left side and posterior aspect of the aorta with an indeterminate endoleak.They physician plans to reline the patient but a date has not been scheduled.The patient is currently asymptomatic.As of this report, there has been no additional patient sequalae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The type iiib endoleak of the right limb complaint is confirmed.The complaint is most likely anatomy related.Procedure related harms for this complaint could not be determined based on the information shared with endologix to review.The final patient status was not reported.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place (b)(6) 2014.The type iiib endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type iiib endoleak events reported for afx devices has been reduced to <0.2%.Root cause investigation was carried out for all afx complaints having an identified failure mode of a type 3b endoleak.Endologix implemented the following corrective actions with the intent of reducing type 3b endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place (b)(6) 2014.The type 3b endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type 3b endoleak events reported for afx devices has been reduced to <0.2%.The review of manufacturing lot confirmed all devices met specifications prior to release.Devices remain implanted in the patient and were not returned, no evaluation completed.The review of manufacturing lot confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
|
| |
|
Search Alerts/Recalls
|