• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the merlin transmitter was not working.It was noted that the patient had a power outage.The prongs were disconnected and burns were noted.The transmitter was replaced.
 
Manufacturer Narrative
Final analysis found burn marks on the main printed circuit board (pcb) and power adapter, and notable burnt damage to the green strips which resulted in power up anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7631844
MDR Text Key112169103
Report Number2017865-2018-09226
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-