SMITH & NEPHEW, INC. SVC REPL, POSITIONER, SPIDER LIMB; ACCESSORIES, OPERATING-ROOM, TABLE (KIT) Back to Search Results
Catalog Number 7210569S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the arm is not locking up, it's slow after the pedal is pressed.A backup device was available to complete the procedure with no delay or patient injuries.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issues were noted.A functional evaluation revealed that the unit would not hold pressure and was low on oil.The complaint was confirmed and the root cause has been determined to be a seal failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL, POSITIONER, SPIDER LIMB
Common Device Name
ACCESSORIES, OPERATING-ROOM, TABLE (KIT)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7632626
Report Number1643264-2018-00492
Device Sequence Number16599056
Product Code FWZ
UDI-Device Identifier03596010647757
UDI-Public(01)03596010647757
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7210569S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Initial Date Received by Manufacturer 06/11/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date06/24/2018
Supplement Report FDA Received Date08/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-