Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2018, the customer reported that the exam was stuck and the report was not available.Merge technical support investigated the customer's allegation by reviewing the merge unity pacs workstation.The audit trail of the exam showed the original exam was read on the same day as the exam date (b)(6) 2018 however the user did not save the report and that there was no save report event on (b)(6) 2018 in the log.There was no data uploaded and therefore the report was not available.The exam was required to be re-read by the physician.The exam was re-read on (b)(6) 2018.While images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm.However, the customer did not allege any harm, serious injury or death as a result of this issue.Reference complaint number: (b)(4).
|