• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 32MM ID US CRMC LINER 48; PROSTHESIS,HIP,SMICONSTRAIND,METL/CERAMC/CERAMC/METL,CMNTDRUCMNTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 32MM ID US CRMC LINER 48; PROSTHESIS,HIP,SMICONSTRAIND,METL/CERAMC/CERAMC/METL,CMNTDRUCMNTD Back to Search Results
Catalog Number 71338948
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Injury (2348)
Date of Event 06/06/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient came into the clinic complaining that she heard a crack or pop while walking.Doctor informed that the r3 ceramic liner is broken.
 
Additional Manufacturer Narrative
The associated device was not returned for evaluation.Our investigation indicates that the reported device was part of field action initiated in 2011.It was identified that the manufacturing process completed for several batches of r3 ceramic liners may have created a lower than expected strength for the liners.It is unclear if the device involved in this complaint failed as a result of this previously identified issue as the device was not returned.A clinical analysis indicated that the primary surgery was performed in (b)(6) 2010.The revision has not been done yet.When the revision reports and medical/clinical documents become available this complaint can be re-assessed.Without the actual device involved, our investigation cannot proceed.No additional actions are being taken at this time.We consider this investigation closed.If additional information is received in the future, this complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 32MM ID US CRMC LINER 48
Common Device Name
PROSTHESIS,HIP,SMICONSTRAIND,METL/CERAMC/CERAMC/METL,CMNTDRUCMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7632976
Report Number1020279-2018-01190
Device Sequence Number11365875
Product Code MRA
UDI-Device Identifier00885556022047
UDI-Public00885556022047
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030022
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2019
Device Catalogue Number71338948
Device Lot Number09FT32646
Initial Date Received by Manufacturer 06/07/2018
Supplement Date Received by Manufacturer06/07/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age59 YR
-
-