• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Occlusion Within Device (1423); Aspiration Issue (2883)
Patient Problem Underdose (2542)
Event Type  Injury  
Manufacturer Narrative
The main component of the system.Other relevant device(s) are: product id: 8731, serial/lot #: (b)(4), ubd: (b)(6) 2006, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a company representative regarding a patient receiving dilaudid 20mg/ml at 1mg/day and bupivacaine 10mg/ml at 0.5mg/day via an implantable pump.The indication for use was non-malignant pain.It was reported that the patient felt he was not receiving drug via his pump.It was unknown if there were any environmental, external or patient factors that may have led or contributed to the issue.The diagnostics and troubleshooting performed was a catheter dye study was attempted, however the catheter was felt to be occluded.The patient was referred to the neurosurgeon for a catheter revision where the patient was scheduled for elective removal and replacement of the intrathecal catheter.The catheter was replaced.The issue was resolved at the time of the report and it was noted that the healthcare provider would have no further information regarding the event.The patient status was noted as alive, no injury and no further complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7633249
MDR Text Key112139748
Report Number3004209178-2018-14179
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530126
UDI-Public00643169530126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2018
Date Device Manufactured03/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-