| Catalog Number |
133FV-14 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
Physician reported "while filling up the te with solution, it was leaking from a hole on the other side.".
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: two openings curved on anterior, one opening curved in the patch and clear fluids inside the device.Leak test and microscopic analysis was performed which identified: two openings punctured curved on anterior and one striated opening in the patch.Based on the device analysis the final assessment is: punctured openings on anterior due to surgical damage like.One striated opening due to surgical damage consist in the use of some surgical tool.
|
| |
|
Event or Problem Description
|
|
Physician reported "while filling up the te with solution, it was leaking from a hole on the other side.".
|
| |
|