| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 85, 82 and 98 mg/dl.The customer's expected fasting blood glucose test result range is 110 - 140 mg/dl.Customer was not using correct testing technique as he was using alcohol to clean his fingers.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification an is stored above the refrigerator in the kitchen.The test strip lot manufacturer's expiration date is 07/24/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history (time not set): (b)(6).
|
| |
|