• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 6935M49
Medical Device Problem Codes Fracture (1260); High impedance (1291); Over-Sensing (1438); Low impedance (2285)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead integrity alert (lia) had triggered due to sensing integrity counter (sic) and a high impedance.It was also noted there was a pacing impedance measurement of 0.A fracture was suspected.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was returned for analysis.In addition to returned product analysis, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product event summary: the partial lead was returned in segments, analyzed.Analysis indicated that the distal conductor of the lead developed a fracture due to flexing while in vivo.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7634559
Report Number2649622-2018-10441
Device Sequence Number13497295
Product Code LWS
UDI-Device Identifier00643169550889
UDI-Public00643169550889
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/13/2017
Device Model Number6935M49
Device Catalogue Number6935M49
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Initial Date Received by Manufacturer 06/02/2018
Supplement Date Received by Manufacturer06/26/2018
08/20/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date07/02/2018
08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVMB1D4 ICD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age9 YR
Patient Weight45
-
-