| Model Number |
6935M49 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Over-Sensing (1438); Low impedance (2285)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead integrity alert (lia) had triggered due to sensing integrity counter (sic) and a high impedance.It was also noted there was a pacing impedance measurement of 0.A fracture was suspected.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was returned for analysis.In addition to returned product analysis, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the partial lead was returned in segments, analyzed.Analysis indicated that the distal conductor of the lead developed a fracture due to flexing while in vivo.If information is provided in the future, a supplemental report will be issued.
|
| |
|