WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/12 HOLES/RIGHT; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
02.118.210 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Impaired Healing (2378); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of device breakage is unknown date approximately one month after implantation.Additional procodes: hwc date of implantation is unknown date in (b)(6) 2017.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history records review has been requested.Code (b)(4) used to capture surgical intervention and device removal.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent removal of hardware and revision to tibial nail and bone graft on (b)(6) 2018, due to a broken plate and nonunion.The patient had an open reduction internal fixation (orif) of the distal tibia in (b)(6) 2017 at an outside facility.The patient fell approximately one month later and broke the plate.One (1) locking compression plate, five (5) unknown cortex screws, one (1) metaphyseal screw, five (5) unknown variable angle locking screws were removed.Fragments were generated from the broken plate but were easily removed.The procedure was completed successfully, with no surgical delay.Patient outcome was reported as stable.Concomitant devices: 3.5 cortex screws (part: unknown, lot: unknown, quantity: 5), 2.7 metaphyseal screw (part: unknown, lot: unknown, quantity: 1), 3.5 va locking screw (part: unknown, lot: unknown, quantity: 1).2.7 va locking screw (part: unknown, lot: unknown, quantity: 4) this report is for a 2.7/3.5mm va-lcp anterolateral distal tibia pl/12 holes/right.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: part: 02.118.210; lot: 7842543; part manufacturing date: november 11, 2014; part expiration date: n/a; manufacturing location: (b)(4) new york.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.7/3.5mm va-lcp anterolateral distal tibia pl/12 holes/right was processed through the normal manufacturing and inspection operations with no non-conformance noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.The review of the raw material device history record revealed this lot met all specifications with no non-conformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|