| Catalog Number |
0006412 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Immucor technical support used a remote electronic connection method on 31may2018 to assess the instrument test well images in question, which appeared as visually positive.There were no errors during processing, and the camera files were acceptable.The immucor technical communication (b)(4) is being used by the lab to visually confirm all negative assay events reported by the echo instrument.Complaints of positive reactions being interpreted as negative by the galileo echo camera algorithm are being corrected under design control project 13-005.(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a customer site reported an unexpectedly positive outcome with anti-d (series 4) monoclonal blend on a galileo echo instrument.
|
| |
|