• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US AVS UNILIF WEDGENOSE VERTEBRAL SPACER 12X25X4-11; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US AVS UNILIF WEDGENOSE VERTEBRAL SPACER 12X25X4-11; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 48395124
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Neuropathy (1983); Pain (1994); Numbness (2415)
Date of Event 12/15/2014
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported that a cage which was implanted on (b)(6) 2014, had slid out from l4/l5 and he will undergo corrective surgery (b)(6) 2018 due to excruciating pain.Patient reported the same type of pain radiating down right leg, neuropathy, and numbness in right foot, which got worse since (b)(6) 2014, when symptoms started.Device will be removed.In addition, patient reported that x-rays show, cage moved about 3/4 inch toward posterior and right and physician stated, "it is partially fused in a bad place.He has to cut it out." patient additionally reported that, "the ct scans show halo around screws which might indicate loosening according to the physician.".
 
Event or Problem Description
The patient reported that a cage which was implanted on (b)(6) 2014, had slid out from l4/l5 and he will undergo corrective surgery (b)(6) 2018 due to excruciating pain.Patient reported the same type of pain radiating down right leg, neuropathy, and numbness in right foot, which got worse since (b)(6) 2014, when symptoms started.Device will be removed.In addition, patient reported that x-rays show, cage moved about 3/4 inch toward posterior and right and physician stated, "it is partially fused in a bad place.He has to cut it out." patient additionally reported that, "the ct scans show halo around screws which might indicate loosening according to the physician.".
 
Additional Manufacturer Narrative
The customer reported event was confirmed upon x-ray images review.Per correspondence with patient, revision surgery was performed and protruding part of implant was removed by the surgeon while the majority part remains inside.No product related issue or surgical technique related issues were observed in the images provided.Cage is observed to be well positioned from the images provided.Device was not returned and lot# was not provided therefore, device inspection, device history review, and complaint history review could not be completed.The cage migration is a known risk and the surgical techniques states that the surgeon must advise the patient of this fact and warn of the potential consequences.Prior to adequate maturation of the fusion mass, implanted spinal instrumentation may need additional help to accommodate full load bearing.Conclusion: the exact root cause of the reported event could not be determined.However some of the plausible contributing factor includes: patient non-fusion, too much instantaneous load, implant under sizing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVS UNILIF WEDGENOSE VERTEBRAL SPACER 12X25X4-11
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7635286
Report Number0009617544-2018-00153
Device Sequence Number13841701
Product Code MAX
UDI-Device Identifier07613252094722
UDI-Public(01)07613252094722
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K093704
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48395124
Device Catalogue Number48395124
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer09/27/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
-
-