• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR) Back to Search Results
Model Number 8000.COM02
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint was reviewed during the retrospective review.An error code, pha 103/104, occurred during surgery, and had to be replaced, hence reportable.This could lead to a prolonged delay.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer ((b)(4)).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
During surgery phaco went lower and lower; changed hp no us power.Error code pha 103/104.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)
Common Device Name
EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7635389
Report Number1222074-2018-00157
Device Sequence Number2044257
Product Code HQC
UDI-Device Identifier08717872019659
UDI-Public08717872019659
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 02/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number8000.COM02
Device Catalogue Number8000.COM02
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2016
Initial Report FDA Received Date06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-