| Model Number |
BLV3220000 |
| Medical Device Problem Code |
Particulates (1451)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sterilization and lot history records were reviewed and found to be acceptable.
|
| |
|
Event or Problem Description
|
|
The user facility reported as the doctor was performing phacoemulsification when a back particle fell into the patients eye.The surgeon was able to remove the particle, and complete the surgery with no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation completed.One small plastic specimen cup was returned.The pack and the components were not returned.Microscopic examination found a dark round particle stuck to the inside wall of the cup.The particle has the appearance of a bead.It is approximately 100 microns in size.This sample was sent to the lab.Further investigation underway.Manufacturing and sterilization records were reviewed and found to be acceptable.Review of the complaint database revealed no other complaints have been submitted against lot #w1448.Further investigation of the particulate is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
The lab analysis indicated the bead-like particle consisted primarily of sulfur, sodium and calcium.The shape and elemental concentrations of the sample suggest that it may be an ion exchange resin bead.Similar beads may be found in the filters of water purification systems used by the hospital or manufacturer.Although we are unable to determine the exact origin of the particle, it is possible the bead may have originated from the b+l water purification system therefore additional analysis will be performed to determine if further action is required.A review of the complaint database revealed no other similar complaints have been reported.The manufacturing and sterilization records were reviewed and found to be acceptable with the device meeting all specifications at the time of release.
|
| |
|