Model Number 97714
Device Problems Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer¿s representative regarding a patient who was implanted with an implantable neurostimulator (ins) for failed back surgery syndrome and spinal pain.It was reported that the patient's battery was overdischarged.They had been in and out of the hospital for the last year and therefore were unable to charge their ins.Additional information was received from the healthcare professional and the manufacturer¿s representative: the overdischarge has not been resolved, the patient declined.Additional information from the health care provider (hcp) via a manufacturer representative notes that the code seen was "0x2618".A trickle charge was performed and the power on reset (por) was cleared.The issue was resolved.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7635487
MDR Text Key112433605
Report Number3004209178-2018-14273
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2018
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Date Manufacturer Received06/19/2018
Date Device Manufactured05/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-