• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number C-UTLM-701J
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pma/510(k) #: k081113.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a triple lumen polyurethane central venous catheter set was used during a research study to retrieve cerebral venous blood samples.The line was in place for approximately 5 hours and samples were taken every 5-10 minutes.The line was flushed frequently with saline after samples were taken.Significant clots within the internal jugular vein were observed and measured via duplex ultrasound.The progression of these clots were tracked and said to dissipate.As reported, the issue of clotting was not noted when different catheters were used.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Correction: the lot number for this device is unknown.Additional information: initially, the customer reported clots formed on the catheter.Additional information provided on 15-aug-2018 from the customer stated the clots were seen via ultrasound along the outside of the device.The device was removed, the patient was given aspirin and a week later the clots were absent during a follow-up echocardiograph.Event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury or product malfunction as the device did not malfunction nor have potential to cause or contribute to patient harm.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER SET
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7635640
Report Number1820334-2018-01923
Device Sequence Number1794750
Product Code DQY
UDI-Device Identifier00827002029378
UDI-Public(01)00827002029378(17)210201(10)8557100
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K081113
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2021
Device Catalogue NumberC-UTLM-701J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/15/2018
Supplement Date Received by Manufacturer08/15/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date09/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-