Model Number 97714
Device Problems Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Battery Problem (2885); Charging Problem (2892)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via manufacturer representative regarding a patient who was implanted with a neurostimulator.It was reported that the device does not hold a charge like it used to.They had not used the device in several months.The implantable neurostimulator (ins) was overdischarged and it was stated that the ins had been overdischarged at least once before.The clinician programmer tried to interrogate the ins but this was unsuccessful because the ins was overdischarged.The patient was educated on a trickle charge and they would be follow-up with on (b)(6) 2018.The issue was not yet resolved.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received on 2018-jul-17 noting that the patient had trouble recharging at home and the device has to be constantly charged.It was too time consuming and hard to keep up with.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from a manufacturer representative (rep) reporting that the cause of the ins not holding a charge was due to the patient not charging it.It was indicated that the ins was able to be brought out of overdischarge.It was indicated that the provided information had been confirmed with the physician.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7635856
MDR Text Key112312365
Report Number3004209178-2018-14280
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2017
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Date Device Manufactured06/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
-
-