MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 'PKG, DEBAKEY FORCEPS'; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 'PKG, DEBAKEY FORCEPS'; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080314
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the product broke inside patient.The broken pieces were successfully removed from patient.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: while pulling mesh through the trocar the hunter bowel snapped and broke while in the patient.The handle and insert were removed the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be excessive force, or rotational force with the jaw under load.Mfg date: manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the product broke inside patient.The broken pieces were successfully removed from patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
'PKG, DEBAKEY FORCEPS'
Common Device Name
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key7636030
Report Number0002936485-2018-00504
Device Sequence Number11211214
Product Code OCZ
UDI-Device Identifier07613327052879
UDI-Public07613327052879
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K973259
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0250080314
Device Lot Number0143178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Initial Date Received by Manufacturer 05/31/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-