| Catalog Number |
0250070601 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that kettle lead started to melt while plugged in to recharge.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: ttle lead melting while plugged in to recharge.Confirmed failure: cosmetic, damaged rear chassis, battery failure past expiration, damaged membrane cover, damaged front panel.Probable root cause: fuse failure.Power transformer/regulator malfunction.Power supply failure.Battery charger malfunction.Incompatible rechargeable battery accessory.User error.
|
| |
|
Event or Problem Description
|
|
It was reported that kettle lead started to melt while plugged in to recharge.
|
| |
|