BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7554
Medical Device Problem Codes Burst Container or Vessel (1074); Retraction Problem (1536); Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty in an upper arm dialysis fistula, the pta balloon allegedly ruptured and was difficult to retract through the sheath, which led to a balloon detachment.The pta balloon and sheath were removed together as one unit.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was returned for evaluation.A visual inspection found a complete burst in the catheter shaft, confirming the investigation for a burst.However, as the balloon was unable to be inflated fully due to the shaft burst, the investigation is inconclusive for the reported sheath-related retraction issue and balloon rupture.The investigation is unconfirmed for the reported detachment as the balloon was returned fully attached to the catheter shaft, and as no other detachments were noted.It is possible that the user perceived the identified catheter shaft burst as the reported balloon rupture, as both failures could contribute to a sheath-related retraction issue.However, the definitive root cause for the reported or identified issues could not be determined based on the available information.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Event or Problem Description
It was reported that during an angioplasty in an upper arm dialysis fistula, the pta balloon allegedly ruptured and was difficult to retract through the sheath, which led to a balloon detachment.The pta balloon and sheath were removed together as one unit.Another pta balloon was used to successfully complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7636304
Report Number2020394-2018-00977
Device Sequence Number2143681
Product Code DQY
UDI-Device Identifier00801741062834
UDI-Public(01)00801741062834
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberCQ7554
Device Catalogue NumberCQ7554
Device Lot NumberREBT1542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer09/11/2018
Initial Report FDA Received Date06/25/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight87
-
-