• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 3120515
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with cervical nerve compression and underwent anterior decompression surgery of the cervical spine.During surgery,the set screws slipped.Hence, they had to be explanted.The procedure was completed successfully with a new product.No adverse event was associated with the screws.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: the screw was returned showing signs of use, hex showing deformation and the threads at the o.D.Of the screw having had the anodization rubbed off from installation.The o.D.Of the screw was checked and meets the print specification.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7636643
Report Number1030489-2018-00900
Device Sequence Number11371483
Product Code KWP
UDI-Device Identifier00613994241146
UDI-Public00613994241146
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063100
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3120515
Device Lot NumberH5393231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Received by Manufacturer 05/29/2018
Supplement Date Received by Manufacturer05/29/2018
09/17/2018
Initial Report FDA Received Date06/26/2018
Supplement Report FDA Received Date07/06/2018
10/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-