• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that the device stalls and it gets hot.No significant delay or patient injury reported.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL MDU CNTRL PWRMX ELITE
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7636650
MDR Text Key112434703
Report Number1643264-2018-00497
Device Sequence Number1
Product Code HAB
UDI-Device Identifier00885556582305
UDI-Public(01)00885556582305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/21/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-