STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 6MM X 20CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035476200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.During visual inspection, the delivery wire was found kinked/bent at the main junction area and the main coil was broken/fractured at the main coil junction area.During functional testing, the main coil was physically detached from the delivery wire.Additional information provided by the customer confirmed that the device was prepared as per device direction for use (dfu) and was in a good condition prior to the procedure.Based on the available information and device analysis, it is probable that procedural factors encountered during the procedure contributed to the reported ¿nv ¿ coil in catheter friction¿ and to the analyzed ¿nv ¿ coil delivery wire kinked/bent¿ and ¿nv - main coil broken/fractured¿.Therefore, as assignable cause of procedural factor was assigned to this investigation.
 
Event or Problem Description
Analysis of the returned device found that the subject coil was broken/fractured.There were no clinical consequences to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 6MM X 20CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7636868
Report Number3008881809-2018-00280
Device Sequence Number2435744
Product Code HCG
UDI-Device Identifier04546540676542
UDI-Public04546540676542
Combination Product (Y/N)N
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Catalogue NumberM0035476200
Device Lot Number18755227
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2018
Initial Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN)
-
-