| Catalog Number |
RTLR180237 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
This is a report of a patient with end stage renal disease (esrd) on continuous cyclic peritoneal dialysis (ccpd) for renal replacement therapy (rrt) who was hospitalized.During follow-up, the peritoneal dialysis registered nurse (pdrn) stated the cause of the hospitalization was unknown, and the patient¿s discharge summary was unavailable.The pdrn stated that the patient was discharged but did not know when.Additional information was requested but unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was returned to the manufacturer for physical evaluation.An external visual inspection was performed with no physical damage noted.A simulated therapy was initiated and completed on the cycler without complication.The weighed fill volumes were found to be within tolerance.The cycler underwent a pressure/vacuum test and was found to meet product specifications.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The cycler performed as designed and an associated cause could not be determined.
|
| |
|
Search Alerts/Recalls
|