MAUDE Adverse Event Report: ST. JUDE MEDICAL BRAIDED SWARTZ¿ INTRODUCER, SL0¿. 8F, 63CM; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRAIDED SWARTZ¿ INTRODUCER, SL0¿. 8F, 63CM; INTRODUCER, CATHETER Back to Search Results
Model Number G407371
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2018
Event Type  malfunction  
Event Description
Related manufacturing ref 3005334138-2018-00177.During a pulmonary vein isolation procedure, air aspirated into the syringe of two introducers.The introducers were replaced and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak is consistent with damage during use.The ifu states, do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAIDED SWARTZ¿ INTRODUCER, SL0¿. 8F, 63CM
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7637411
MDR Text Key112310223
Report Number3005334138-2018-00178
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Model NumberG407371
Device Lot Number6277004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BRAIDED SWARTZ¿ INTRODUCER
-
-