MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. EP-WORKMATE¿ AND WORKMATE¿ CLARIS¿ EP-4¿ TOUCHSCREEN COMPUTER; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. EP-WORKMATE¿ AND WORKMATE¿ CLARIS¿ EP-4¿ TOUCHSCREEN COMPUTER; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number H21911
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/29/2018
Type of Reportable Event Malfunction
Event or Problem Description
During device preparation, smoke emitted from the back of the touchscreen.There was no patient involved, therefore there were no adverse patient consequences.
 
Additional Manufacturer Narrative
The results of the investigation confirmed electrical overstress (eos) damage on the hard disk drive power connector.The smoke observed in the field was consistent with the electrical overstress damage observed on the hard disk drive power connector.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EP-WORKMATE¿ AND WORKMATE¿ CLARIS¿ EP-4¿ TOUCHSCREEN COMPUTER
Common Device Name
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key7637483
Report Number2184149-2018-00091
Device Sequence Number19461085
Product Code DQK
UDI-Device Identifier05414734219112
UDI-Public05414734219112
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K132073
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH21911
Device Lot Number6504202
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/29/2018
Supplement Date Received by Manufacturer07/27/2018
Initial Report FDA Received Date06/26/2018
Supplement Report FDA Received Date08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-