| Model Number |
73875-50 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Information provided by the international customer indicated they believe the event was due to a bad connection between the instrument & polyaxial screw.Visual inspection appears to confirm this report.An evaluation of the returned polyaxial screw found one side of the tulip has been completely broken/detached from the remainder of the implant.Additionally, both the set screw & polyaxial screw appear to have gouges and markings indicative of cross threading.It appears likely that the anti-torque may not have been completely seated and when forces were placed on the screw during the final tightening process.
|
| |
|
Event or Problem Description
|
|
An international customer ((b)(6)) reported that the cannulated screw located at the iliac broke during the final tightening process.The screw was removed and replaced without issue.
|
| |
|