Batch review performed on 22 june 2018.Lot: 103102; (b)(4) items manufactured and released on 30 november 2010.Expiration date: 2015-10-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Clinical evaluation performed by medical affairs director on 25 june 2018: revision of hybrid tha in a (b)(6) woman after 7 years.The radiographs show a cement deficiency in the proximal medial portion and probably discontinuity distally.These conditions probably led to distal load transfer and the subsequent radiographs show the consequences of this mechanical pattern.The cause of issue may lie in the procedure of cement application/implant insertion.There is no reason to suspect that a malfunctioning device originated the adverse event.The purported hypothesis of stem corrosion is not supported by evidence and it is confirmed to be irrelevant to the case.Visual inspection performed by r&d hip manager on 08 june 2018 and 20 june 2018: the stem shows large areas of dark staining, that had been suggested to be signs of corrosion.These have been delicately wiped with soft tissues wet with alcohol and were easily removed, therefore clearly showing that they were nothing but organic deposits, most probably blood, a rather common finding on a surgical implant.Said removal treatment was executed on a minimal part of the stain so that it can be later verified by third parties.The distal portion of the stem looks opaque and was again checked for possibility of corrosion.No evidence of corrosion was found, and the most likely explanation for the opaque look is that the stem was implanted in a way that allowed repeated movement within the cement mantle, causing the surface finish to be smudged over the years.The ceramic head was also visually analysed.The negative 12/14 conical seating was found to be regular, with metal transfer signs evenly distributed, no evidence of abnormal stress concentration.Very evident are two large black stripes on the articulating surface, clear signs of metal transfer.Our hypothesis is that metal particles have been trapped in the articulating space and were spread over a very thin layer on the head surface.They are most likely derived from either instruments or implants that may have been damaged at the time of surgery.Such findings on the head contribute to the development of abnormal wear on the polyethylene insert because they act as a third body and macroscopically damage the surface finish of the smooth head.Warnings to check for these situations are given in the device instructions for use and in the surgical technique.
|