| Catalog Number |
8400.25A01 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/28/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(6) 201 the investigation was done on the returned needle.The needle only had a few scratches because the hand piece and the material that came off during surgery where not returned we where not able to determine the point of cause.Surgeon is expecting a medical intervention after his diagnosis on a week because of the broken particle.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
During surgery, the surgeon realized that there was a loose tip of metal on the tip of a phaco which had not broken yet.This was while the phaco was inside the patient`s eye and doctor had to take phaco out of patient`s eye.As a result of this event, there was a tear in the posterior capsule.According to the surgeon, it appears as though a part from the inside of the cutter is what broke off, internally from the tip.The surgeon wil be seeing the patient in a few weeks, only then will he know how the patient was affected.
|
| |
|
Additional Manufacturer Narrative
|
|
This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre b.V.).Details of this activity were discussed with cdrh office of compliance (b)(6) during a tele-conference on 05/31/201.The investigation was done on the returned needle.The needle only had a few scratches because the hand piece and the material that came off during surgery where not returned we where not able to determine the point of cause.Surgeon is expecting a medical intervention after his diagnosis on a week because of the broken particle.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Search Alerts/Recalls
|