Same case as mdr id: 2134265-2018-05723.It was reported that the burr became stuck on the rotawire.The 90% stenosed target lesion was located in the moderately tortuous, severely calcified and 3.7mm in diameter proximal right coronary artery.A rotawire¿ and a 2.25mm rotalink¿ plus were selected for use.During the procedure, a 1.75mm burr was initially used due to the high stenosis rate and severe calcification.When the vessel diameter was measured by ivus, the size was 3.7mm; thus the burr was exchanged with a bigger size of 2.25mm.When the 2.25mm rotaburr was inserted into a non-bsc guide catheter, severe resistance was encountered.The physician attempted to further advance the burr through dynaglide mode; however, it was observed that the burr became stuck on the rotawire.The procedure was completed with a 2.15mm rotaburr and another of the same rotawire to ensure patient safety.No patient complications were reported.
|