• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GEN II CEMENTED TIBIAL B/PLATE SZ 1 LEFT; PRSTHSIS,KNEE,PATLLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD, POLYMER/METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GEN II CEMENTED TIBIAL B/PLATE SZ 1 LEFT; PRSTHSIS,KNEE,PATLLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD, POLYMER/METAL/POLYMER Back to Search Results
Catalog Number 71420160
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Information (3190)
Date of Event 04/18/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to the legion insert being disassociated from the baseplate.
 
Additional Manufacturer Narrative
The associated legion ps high flexion insert and genesis ii tibial baseplate were returned and evaluated.A lab analysis conducted during this investigation noted that burnishing and scratches on the tibia baseplate and insert where they meet are likely caused by repeated motion between them.The locking mechanism on the insert is deformed, likely due to removal.The exact cause of dislocation could not be determined, based on the results of this investigation.A dimensional evaluation was attempted on the insert; the damage/deformation at several features of the poly insert would not allow for accurate measurement.The tibia baseplate measures to be within specification.Our investigation including a review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II CEMENTED TIBIAL B/PLATE SZ 1 LEFT
Common Device Name
PRSTHSIS,KNEE,PATLLOFMROTIBIAL,SMI-CNSTRAIND,CMNTD, POLYMER/METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7638373
Report Number1020279-2018-01203
Device Sequence Number2376359
Product Code JWH
UDI-Device Identifier03596010206237
UDI-Public03596010206237
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2024
Device Catalogue Number71420160
Device Lot Number14HT45030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer10/28/2018
Initial Report FDA Received Date06/26/2018
Supplement Report FDA Received Date11/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-