• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
The customer reported the battery pca had bent pins.The device was not in use on a patient.
 
Manufacturer Narrative
One battery pca was returned for failure analysis.The reported problem was verified during visual inspection.J1 pin 8 was found to be broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key7638622
MDR Text Key112447558
Report Number1218950-2018-05241
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-