| Catalog Number |
02K91-39 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.An evaluation is in process.A final report will be submitted when the evaluation is complete.This report is being filed on an international product, architect ca19-9 list (b)(4) that has a similar product distributed in the us, list number (b)(4).
|
| |
|
Event or Problem Description
|
|
The following data was provided from the architect sid (b)(6) initial 1865.3 u/ml.Which was compared to a result from another method: 500 u/ml.There is no information on the patients clinical history.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling.No adverse trend was identified for the customer's issue.No return patient sample was available.Historical performance of the reagent lot was evaluated using world wide data.The patient data was analyzed and within established control limits.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|