| Catalog Number |
222286 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/02/2013
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident as related to the reported problem, and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed one other dissimilar complaint for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was initially reported one anchor broke and one came out.The procedure was successfully finished with a like device.Additional complaint event information was received from our affiliate on may 15, 2013.The patient had normal bone quality, a healix awl was used in the procedure, and two (2) arthrex fiber wire sutures were being used with each anchor.There was no audible or tactile indication of failure with either anchor.The surgeon was not using the product for the first time.For the anchor which pulled out during the procedure, it pulled out during verification of the repair strength.It is unknown what type of force was applied, and unknown if the tendon side or free ends of suture were tested to verify the repair when the anchor pulled out.For the anchor which broke during the procedure, the anchor was not inserted off-axis and not used as a lever.The anchor broke while pulling suture through the system.It is unknown how long the suture limbs were, if they were of equal length, or if they had become tangled.Please see also medwatch report 1221934-2018-50870.
|
| |
|
Search Alerts/Recalls
|
|
|