Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results of 46 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 120 mg/dl.At the time of the call, the customer felt well and did not report any symptoms.Customer stated when she had obtained the result of 46 mg/dl she was feeling shaky.Customer had some juice and glucose tabs, tested again and obtained 61 mg/dl non-fasting.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.Customer also stated she was comparing the truemetrix meter with her truetrack meter; actual meter to meter comparison results were not provided.Customer declined to perform a back to back blood test during the call on (b)(6) 2018.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 09/28/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1 :61mg/dl date:(b)(6) 2018 time:09:32am non-fasting, result 2 :46mg/dl date:(b)(6) 2018 time:09:23am fasting, result 3 :104mg/dl date:(b)(6) 2018 time:07:09am non-fasting, result 4 :108mg/dl date:(b)(6) 2018 time:07:07am fasting, result 5 :69mg/dl date:(b)(6) 2018 time:06:29am fasting.Customer requested we send out control solution for her to test the meter.
|