MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA MRI AF XT VR ICD; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA MRI AF XT VR ICD; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVFB1D4
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 06/04/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was shocked by the implantable cardioverter defibrillator (icd), possibly due to atrial fibrillation (af) with rapid ventricular response.The device remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The device was later explanted and replaced when the patient received a crt-d device.
 
Manufacturer Narrative
The device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA MRI AF XT VR ICD
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7639360
MDR Text Key112383718
Report Number3004209178-2018-14398
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00643169929913
UDI-Public00643169929913
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2019
Device Model NumberDVFB1D4
Device Catalogue NumberDVFB1D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/04/2018
Initial Date FDA Received06/26/2018
Supplement Dates Manufacturer Received07/30/2018
09/11/2018
Supplement Dates FDA Received08/10/2018
09/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6935M62 LEAD
Patient Outcome(s) Life Threatening;
Patient Age74 YR
-
-