| Catalog Number |
07D70-31 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely elevated magnesium results on the architect c16000 analyzer.The following data was provided (mg/dl): patient 1 initial 7.5, repeat 2.1; patient 2 initial 6.0, repeat 2.3.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
On 4july2018, the suspect medical device was updated to the architect analyzer.See 1628664-2018-00256 for any further investigation and information.
|
| |
|