| Model Number |
DCJLR-A |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
06/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: actual device not evaluated.Dhr review was not performed because the complaint is unrelated to product performance or packaging.Results: no results available since no evaluation performed.Conclusion: human factors issue.Training deficiency.Device not returned.Accriva diagnostics has requested all data required for form 3500a.
|
| |
|
Event or Problem Description
|
|
Healthcare professional called to report an end-user injury that occurred during dispensing a directcheck act-lr abnormal liquid quality control.Lyophilized control material is packaged in a glass ampule inside a crushable plastic dropper vial containing diluent.Each directcheck control includes a set of protective sleeves intended to safeguard end-users from potential injury while handling the control.The end-user utilized the protective sleeve when he crushed the ampule to mix the reagent and diluent.He removed the protective sleeve after mixing and squeezed the vial to dispense the reconstituted control.When doing so, a shard of the glass ampuole penetrated the vial and caused a small cut to the right middle finger.The end-user washed the finger with soap and water, then affixed a band aid.Severe bleeding or other medical complications were recorded.
|
| |
|
Event or Problem Description
|
|
Follow-up #1.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|