• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT; MITRACLIP DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT; MITRACLIP DELIVERY SYSTEM Back to Search Results
Catalog Number CDS0502
Device Problems Failure To Adhere Or Bond (1031); Difficult to Remove (1528)
Patient Problem Tissue Damage (2104)
Event Date 06/11/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report tissue damage and mitral valve replacement surgery.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The clip delivery system (cds) was advanced to the leaflets.The patient had a flail and leaflet grasping was difficult, but once grasped, the gradient was noted to be high.During removal of the cds, the clip became caught in the chordae and a chordal rupture and flail was noted.Due to the high gradient, the procedure was aborted.On (b)(6) 2018, the patient was taken to surgery for mitral valve replacement.Post-surgery, the patient was stable.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The reported patient effect of mitral valve injury (tissue damage) as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported difficulty grasping appears to be related to patient morphology/pathology and due to leaflet flail.The reported difficulty removing the clip from the anatomy (chordae) appears to be due to a combination of patient anatomy and procedural circumstances of difficulty grasping.The tissue damage was likely a result of the clip becoming caught in the chordae (difficult to remove) and therefore, related to procedural circumstances.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacturing or labeling of the device.
 
Event Description
Subsequent to the initial 30-day medwatch report, it was noted that during grasping, the clip became caught in the chordae.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM CLIP DELIVERY SYSTEM NT
Type of Device
MITRACLIP DELIVERY SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key7640178
MDR Text Key112413600
Report Number2024168-2018-05066
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/06/2019
Device Catalogue NumberCDS0502
Device Lot Number80105U274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2018
Initial Date FDA Received06/26/2018
Supplement Dates Manufacturer Received08/22/2018
Supplement Dates FDA Received08/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Disability;
-
-