MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SOFSILK; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SOFSILK; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170002
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Human-Device Interface Problem (2949)
Health Effect - Clinical Codes Tissue Damage (2104); No Known Impact Or Consequence To Patient (2692)
Date of Event 06/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic gastric bypass, the needle was difficult to toggle and would not slide through the tissue.So they physically had to slide the tissue off of the needle which caused a larger than normal hole in the tissue from the needle.4 devices were causing issues during the case.They kept trying to get a new device and the new ones were still causing issues.There was sticking when toggling the device, loading and unloading, the device would also click loudly on some when being toggled and the needle was loose on some as well.More devices were opened to complete the case.The tissue was stretched due to struggling to get the needle out of the tissue so the hole was larger than normal.There was no patient injury.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFSILK
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7641343
Report Number9612501-2018-01200
Device Sequence Number2512722
Product Code OCW
UDI-Device Identifier20884521126852
UDI-Public20884521126852
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number170002
Device Catalogue Number170002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer09/10/2018
Initial Report FDA Received Date06/27/2018
Supplement Report FDA Received Date10/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-