MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. 45MM ARTICNG LNR CUTR/6 ROW BL; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. 45MM ARTICNG LNR CUTR/6 ROW BL; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number 6TB45
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch number: n54h01.Investigation summary: the analysis results found that the 6tb45 device was received with no apparent damaged and with a tr45w cartridge loaded on the device.The returned reload was partially fired 1/10.The returned device was tested for functionality with a test reload and it fired, cut, and all the staples formed as intended.The device fired without any difficulties, the staple line was complete and the staples were noted to have the proper b-formed shape.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lock out.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event or Problem Description
It was reported that the device could not fire.During the radical resection of rectal cancer, could not fire.Another device was used to complete the surgery.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
45MM ARTICNG LNR CUTR/6 ROW BL
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7641456
Report Number3005075853-2018-10951
Device Sequence Number1556229
Product Code GDW
UDI-Device Identifier20705036000874
UDI-Public20705036000874
Combination Product (Y/N)N
PMA/510(K) Number
K070887
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/22/2021
Device Catalogue Number6TB45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2018
Initial Report FDA Received Date06/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-