MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
ETLW1628C93E |
| Medical Device Problem Codes |
Leak/Splash (1354); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Rupture (2208)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm of unknown size.It was reported that, approximately 2 years post index procedure, the patient presented emergently with a ruptured abdominal aortic aneurysm.Ct on this date showed left iliac limb migration from the left iliac into the aneurysmal sac with a type ib endoleak.The patient received treatment; a non medtronic limb was extended to the external iliac and the internal iliac was occluded with a coil to resolve the event.Per the physician, the cause of the event cannot be determined.No additional clinical sequelae were reported, and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received: it was reported that the patient contacted technical services to report his presentation to hospital for diagnosis of an unknown endoleak.The patient reported the stent graft was explanted and replaced with an unknown manufacturers stent graft.The patient did not know the date of the emergency explant, the hospital where the procedure took place or the physician that carried the procedure.The physician has confirmed that an explant procedure has not taken place and the last procedure completed with this patient was as previously stated, when the patient received intervention for the type ib endoleak and migration.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Film evaluation summary: the exact cause of the events could not be determined; however, it is likely that there has been disease progression.Films prior to implant were not available, the anatomy at implant is unknown, and earlier post-implant films were not provided for comparison.Ct¿s returned from 27 months post-implant revealed that an endurant iis bifurcate, endurant flared limb, and a non-medtronic stent graft system had been implanted in the patient.The max diameter aaa measured 100mm and both common iliacs were also aneurysmal.The endurant bifurcate was positioned 10 -15mm below the lowest right renal artery within the top of the aaa sac, and there was lack of proximal stent graft apposition within the aorta and a proximal type i endoleak was visible.The bifurcate proximal od measured ~37mm, and the aorta at that level was ~50mm in diameter.The infrarenal neck and aortic body were angulated at the gate ring ~80 deg a-p and 30 deg rt-lt; relative to the distal aorta.The films also revealed that the bifurcate ipsi limb on the left had been extended with an endurant flared limb; however, the left limb had evidently pulled out of the left common iliac and was currently positioned within the bottom of the aaa sac.The distal od of the left flared limb measured 30mm, and the diameter at the origin of the left common iliac artery measured ~30mm.A non-medtronic limb was seen having been placed within the left limbs, from the bifurcate flow divider extending distally into the left external iliac; bypassing the left internal which had been coiled.It appears likely that disease progression with iliac artery dilatation may have led to the left limb pulling out of the lcia and distal type i endoleak, which may have contributed to the aaa rupture.Of note, a non-medtronic limb had been placed into the bifurcate gate and extended distally to the right common iliac.An additional non-medtronic bifurcate had been implanted into the right/contra limb with the ipsi limb extending into the right external and the additional contra limb positioned into the right internal iliac artery.It is therefore possible that the other manufacturer¿s devices implanted on the right side may have also contributed to the aaa rupture.If information is provided in the future, a supplemental report will be issued.
|
| |
|