• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MX-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Deflation; patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Healthcare professional reported a deflation a tissue expander.The expansion process was completed.Device has been explanted.This record is for an unknown side.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified more than three curved openings and black particles in device outer surface.A leak test was performed which identified openings.A microscopic analysis was performed which identified more than three curved striated openings.Based on the device analysis the final assessment is: more than three curved striated openings assessed as surgical damage.
 
Event or Problem Description
Healthcare professional reported a deflation of a tissue expander.The expansion process was completed.Device has been explanted.This record is for an unknown side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7641975
Report Number9617229-2018-04640
Device Sequence Number19461120
Product Code LCJ
UDI-Device Identifier10888628004504
UDI-Public10888628004504
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2021
Device Catalogue Number133MX-15-T
Device Lot Number3126597
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2018
Initial Date Received by Manufacturer 06/04/2018
Supplement Date Received by Manufacturer07/16/2018
Initial Report FDA Received Date06/27/2018
Supplement Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-